The Vayyar Care device must be installed on the wall of the room it will be monitoring, in the center of the room, and as far from metal objects as possible.
The device must be placed at a height of 5 feet (1.5 meters) from the floor to the center of the device, oriented straight, and flat on the wall (not twisted or tilted) using the Velcro strips on the back of the device.
The device comes with two strips already stuck to the back of the device and two additional strips attached to them. Simply peel the film off the back of the strips and stick the device to the wall at the correct height of five feet (1.5 meters).
The device needs to be plugged into an electrical outlet and must have a clear view of all parts of the room. It must not be positioned behind a TV, for example.
After the Vayyar Care device is installed and plugged in, follow the instructions in the Vayyar Care app to pair the device, set the perimeter of the room, and perform a test fall.